Damage to Polymer Coatings on Microguidewire Tips through Shaping and J-wiring for Optimizing Flow Diverter Deployment: A Scanning Electron Microscopy Study

Article information

Neurointervention. 2026;21(1):44-52
Publication date (electronic) : 2026 January 12
doi : https://doi.org/10.5469/neuroint.2025.01130
1Department of Radiology, Copenhagen University Hospital - Rigshospitalet, Copenhagen, Denmark
2Department of Chemistry, Technical University of Denmark, Kongens Lyngby, Denmark
3Department of Radiology, Baylor College of Medicine, Houston, TX, USA
Correspondence to: Goetz Benndorf, MD, PhD Department of Radiology, Baylor College of Medicine, 1 Baylor Plz, Houston, TX 77030, USA E-mail: benndorf@bcm.edu
This work was presented at the 21st Annual Meeting of the Society of Neurointerventional Surgery in Colorado Springs, CO, USA (July 2024).
Received 2025 November 29; Revised 2025 December 22; Accepted 2025 December 26.

Abstract

Achieving adequate wall apposition is a crucial technical goal when deploying flow diverters to treat wide-neck cerebral aneurysms. The socalled J-wiring technique is a common method used to optimize flow diverter wall apposition. However, the frictional interaction between the shaping device and the microguidewire tip during the formation of the J-loop, as well as the interaction between the J-loop and the flow diverter during J-wiring, may potentially cause damage to the guidewire’s coating. Three frequently used guidewires were tested in vitro in a silicone aneurysm model. Manual J-shaping of guidewire tips, along with the J-wiring technique (including J-shaping), caused damage to the surface coating of guidewires, as observed by scanning electron microscopy. Therefore, both mechanisms may contribute to the generation of polymer micro-fragments in patients treated with flow diversion.

INTRODUCTION

Delayed onset of non-ischemic cerebral enhancing (NICE) lesions following endovascular procedures is a rare phenomenon with a distinctive MRI pattern, characterized by multiple punctate, annular, or nodular contrast enhancements in the brain parenchyma downstream from the vascular territory accessed during endovascular therapy (EVT) [1]. These enhancing foci usually have a cortical, subcortical, or leptomeningeal distribution and a perifocal edema [1]. They are thought to represent inflammatory foreign body reactions to cerebral emboli of polymer micro-fragments originating from polymer-coated endovascular devices [2].

In the literature, NICE lesions have mainly been reported following endovascular treatment of cerebral aneurysms [1,3-6]. While benchtop testing of endovascular devices has identified some causes of coating damage and micro-fragment generation [7-13], the range of mechanisms that can damage polymer-coated devices during aneurysm treatment is not fully understood. The so-called J-wiring technique is widely used in patients treated by flow diversion to improve flow diverter wall apposition [14]. However, this technique may potentially damage polymer coatings due to frictional interaction between (1) the shaping device and microguidewire tip when creating the 180° J-loop and (2) the interaction between the J-loop and flow diverter during J-wiring [15]. In this study, we examined 3 microguidewires for coating damage using scanning electron microscopy (SEM) after applying the J-wiring technique to optimize flow diverter wall apposition in an in vitro aneurysm model.

MATERIALS AND METHODS

Experimental Model

A 4.5x12 mm PipelineTM Vantage Embolization Device with Shield TechnologyTM (Medtronic) was deployed using a PhenomTM 27 Catheter (Medtronic) in a silicone aneurysm model. The model consisted of a 6-mm aneurysm on a curved segment with an intraluminal diameter of 3.8 mm (Codman Neuro). The flow diverter remained stable and fully expanded during guidewire testing.

Guidewire Testing

Three commonly used commercially available 0.014″ guidewires, including the AvigoTM (Medtronic), Synchro® (Stryker), and TraxcessTM (Terumo Neuro), were tested. Four unused samples of each guidewire were obtained: (1) 1 of each guidewire served as a control and was neither shaped nor used for the J-wiring technique; (2) 1 of each guidewire was shaped into a J-loop, but not used for the J-wiring technique; and (3) 2 of each guidewire had their distal ends shaped to a J-loop and were used to simulate the J-wiring technique in the aneurysm model. All devices were used following the manufacturer’s instructions. The guidewire tips were shaped into a 180° curve (J-loop) using a standard metal shaping tool. The J-wiring technique was simulated by inserting the guidewire into the aneurysm model via the PhenomTM 27 and moving its tip back and forth 3 times inside the flow diverter under fluoroscopic guidance (Fig. 1). The shaped guidewire tips were moved over the full length of the flow diverter between its proximal and distal ends. The distal 10 mm of all guidewires was cut and stored in dry test tubes.

Fig. 1.

Radiograph of the silicone vascular model with a PipelineTM Vantage flow diverter (Medtronic) deployed in a curved segment with a 6-mm aneurysm. The J-wiring technique was performed by shaping the tips of the AvigoTM (Medtronic) (shown), Synchro® (Stryker), and TraxcessTM (Terumo Neuro) guidewires into 180° J-loops and moving the J-loop back and forth 3 times inside the flow diverter. The curved arrow indicates the back-and-forth movement of the guidewire tips during the J-wiring technique. The guidewires were moved between the proximal and distal ends (full length) of the flow diverter. Guidewire tips were cut immediately after withdrawal from the model and subsequently imaged by scanning electron microscopy.

Scanning Electron Microscopy

All guidewire tips were examined using SEM (FEI Quanta 3D FEG; Thermo Fisher Scientific). Control wires were flushed with demineralized water, cut directly after unpackaging, and dried before mounting. All samples were mounted on 12 mm aluminum stubs and sputter-coated with 6 nm gold particles before SEM analysis (ACE200; Leica Microsystems). Because of the mounting method, only 1 side of each guidewire tip could be imaged. The entire length of each guidewire tip was inspected for coating damage by SEM, and high-resolution images of areas with the most significant damage were captured. Images were evaluated for coating damage on both the inner (concave) and outer (convex) curves of the guidewire’s J-loop. Coating damage was identified as defects or irregularities in the guidewire surface. Material on the guidewire surface was not assessed, as its origin could not be determined; although it resembled micro-fragments, it could also be contamination with fiber or dust particles. Results were described qualitatively and supported by representative SEM images.

RESULTS

AvigoTM Guidewire

Representative images are shown in Fig. 2. The control guidewire remained intact. J-shaping alone caused continuous coating damage to the inner and outer curves of the guidewire, leaving a rough and irregular surface. The J-wiring technique (including J-shaping) caused continuous coating damage to the inner and outer curves of both guidewires.

Fig. 2.

Scanning electron micrographs of the AvigoTM (Medtronic) guidewire tip. (A, B) Unshaped control: intact coating with a few small irregularities. (C, D) J-shaping alone: continuous coating damage on both the inner and outer curves of the guidewire (arrows). (E, F) The J-wiring technique: continuous coating damage on both the inner and outer curves of the guidewire (arrows).

Synchro® Guidewire

Representative images are shown in Fig. 3. The control guidewire remained intact. J-shaping alone did not cause any coating damage. The J-wiring technique (including J-shaping) resulted in minor coating damage to the inner and outer curves of both guidewires, spanning multiple consecutive levels between gaps in the hypotube, which exposed a slightly irregular surface. In one area, delamination of the probably deeper irregular surface led to a partially attached micro-fragment measuring 9 μm in length (Fig. 3D). The guidewires exhibited a few focal areas with different types of coating damage, including long, linear scratches (Fig. 3E) and localized areas where the coating was “dragged” along the hypotube (Fig. 3F). The latter type of damage was found on the inner curve of the J-loop and was caused by a force applied in the direction from proximal to distal (right to left in Fig. 3F), most likely during the shaping maneuver.

Fig. 3.

Scanning electron micrographs of the Synchro® (Stryker) guidewire tip (polymer-coated hypotube). (A, B) Unshaped control: the coating is intact. The guidewire was also intact after J-shaping alone (not shown). (C–F) The J-wiring technique: slight coating damage were found on both the outer (arrows in C and D) and inner (not shown) curves of the guidewire on multiple consecutive levels of the hypotube, exposing an irregular surface. A partially attached micro-fragment probably coming off the deeper irregular surface (arrowheads in D). Other types of coating damage that spanned multiple gaps in the hypotube include long, linear scratches (arrows in E) and focal areas of coating being dragged along the hypotube (arrows in F).

TraxcessTM Guidewire

Representative images are shown in Fig. 4. The control wire had one focal area with coating damage, but was otherwise intact. J-shaping alone caused coating damage to the inner and outer curves of the guidewire, spanning multiple consecutive levels of the coiled wire, exposing longitudinal grooves in the wire surface, possibly from the wire drawing process. The J-wiring technique (including J-shaping) caused coating damage to the inner and outer curves of both guidewires, spanning multiple consecutive levels of the coiled wire, which seemed to be more extensive than after J-shaping alone. The coating was mostly damaged on the outer crest (central part) of the coiled wire, while it was intact in the valleys (gaps) between the coiled wire.

Fig. 4.

Scanning electron micrographs of the TraxcessTM (Terumo Neuro) guidewire tip (polymer-coated coiled wire). (A, B) Unshaped control: the guidewire is intact apart from a single focal area of coating damage (arrows). A few irregularities in the coating in the gaps between the coiled wire are observed. (C, D) J-shaping alone: coating damage on both the inner and outer curves of the guidewire (arrows) on multiple consecutive levels of the coiled wire. (E, F) The J-wiring technique: coating damage on both the inner and outer curves of the guidewire (arrows) on multiple consecutive levels of the coiled wire, most extensive on the outer curve. A close-up of the coating damage in (F) shows multiple longitudinal grooves in the coiled wire. The coating damage mainly involves the central part of the coiled wire, while the coating in the gaps between the coiled wire remains intact.

DISCUSSION

NICE lesions are thought to be granulomatous foreign-body reactions to cerebral emboli of hydrophilic polymer micro-fragments. They represent a rare complication to neuroendovascular therapy that sometimes may require long-term immunosuppressive treatment. NICE lesions develop following 0.05–2.30% of procedures in patients undergoing EVT for cerebral aneurysms and have been reported after simple coiling, stent/balloon-assisted coiling, and flow diversion [1,3-6]. Polymer coatings may detach from endovascular devices either through time-dependent chemical degradation or mechanical damage [16]. The continuous degradation of polymer coatings in contact with saline or blood increases over time and may result in the shredding of polymer fragments within 15 minutes [8,9]. Mechanical damage to coatings occurs when polymer-coated devices forcefully interact with each other, the vasculature, or with tools used for endovascular procedures. This can result in the breakage of chemical bonds within the polymer coating or lead to the separation of the coating from the deeper structure of the device [16]. Avoiding mechanical coating damage seems the most feasible way to reduce the risk of NICE lesions during endovascular procedures.

In this small study, we show that J-shaping alone and the J-wiring technique (including J-shaping) cause considerable mechanical damage to polymer coatings on guidewire tips. Firstly, the manual shaping of the guidewire tip into a 180° J-loop caused damage to both the outer and inner curves of the J-loop in 2 of 3 guidewires. While some coating damage on the inner curve of the guidewire tip was expected after shaping due to the mechanical force from the metal shaping tool, we were surprised to find coating damage on the outer curve, which was only touched by the glove during the shaping maneuver. More tests are needed to determine whether coating damage occurs when ‘shapeable tips’ are curved to less than a 180° J-loop, which is commonly performed during various endovascular procedures. Guidewires are usually inserted directly after shaping, and may thus pose a potential risk of inadvertently introducing polymer micro-fragments into the cerebral circulation if shaping causes coating damage. Therefore, it seems advisable to consider using preshaped guidewires when possible or to flush guidewire tips with saline after a shaping maneuver to remove potential coating debris.

Secondly, the J-wiring technique (including J-shaping) to optimize flow diverter wall apposition caused damage to both the outer and inner curves of the J-loop on all 3 guidewires. Since J-shaping alone caused coating damage, we cannot establish to what degree the J-wiring step itself contributed to the overall damage. However, coating damage appeared more extensive after the J-wiring technique compared to J-shaping alone in 2 of 3 guidewires, suggesting that additional damage may be caused by the backand- forth movement of the J-loop within the flow diverter (J-wiring step). It is conceivable that a damaged coating integrity during the initial manual J-shaping increases the risk of further damage, thereby generating polymer micro-fragments during the J-wiring step. Importantly, this is of particular concern because this additional abrasion and shedding would migrate directly into the cerebral circulation. The contribution of the J-wiring step to coating damage could more precisely be explored in future studies by analyzing water samples drawn from the aneurysm model for micro-fragments.

The vulnerability of polymer-coated devices to coating damage can be attributed to both mechanical design and coating composition [17]. We observed differences in the overall pattern of damage among guidewires, primarily determined by the guidewire tip design. On the smooth cylindrical AvigoTM tip, frictional stresses were distributed continuously along the surface, resulting in continuous abrasions. For the Synchro® tip, the polymer coating is applied over a hypotube with small gaps that act as “stress-relief” points, so that damage appeared as short, segmented rather than continuous abrasions. For the TraxcessTM tip with a coiled design, contact occurred mainly on the outer coil crests, concentrating damage in these regions while leaving the valleys largely intact. Therefore, the wire tip architecture determines where stresses build up during shaping and manipulation, which may also have implications for the size and shape of generated micro-fragments. While longer coating fragments could potentially peel off the AvigoTM guidewire tip, the length of fragments coming off the TraxcessTM and Synchro® guidewire tips would probably be shorter than the diameter of the distal coiled wire (~65 μm) and the distance between gaps in the hypotube (~70 μm). Previous clinical and in vitro studies have reported a large variability in the size of micro-fragments [15,18]. Furthermore, we found indications of damage involving the superficial coating and probably also the deeper structure of the Synchro® guidewire after the J-wiring technique, highlighting that micro-fragments of various materials may be generated from the same guidewire [15,19].

The chemistry of a polymer coating also influences how the material fails under stress. To our knowledge, the exact chemical composition of polymer coatings on the AvigoTM, TraxcessTM, and Synchro® tips is not publicly disclosed. However, hydrophilic coatings, such as polyvinylpyrrolidone-based coatings, are soft and compliant, allowing them to deform and wear away gradually rather than fracture [20]. Polytetrafluoroethylene-based coatings are relatively brittle and exhibit low surface adhesion, so they tend to crack or flake when the interface with the metal substrate is weakened [21]. Polyurethane- or polyamide-based coatings are more elastic and resistant to crack propagation, so damage is confined to small torn regions at points of highest stress [22]. Thus, these commonly used polymer coatings differ in the energy required for a crack to propagate or for the coating to separate from the substrate, which may explain why some coatings fail by brittle flaking and others by gradual abrasion.

Our experiment provides evidence for a new mechanism contributing to coating damage on microguidewires during flow diversion, adding to the list of routinely performed procedural steps during EVT that may result in coating damage. Previously, backloading of microguidewires through insertion tools has been associated with coating damage and micro-fragmentation [15,19]. In addition, coaxial manipulation of combinations of microcatheters, catheters, and sheaths has been associated with coating damage in benchtop tests [10-12]. Tight-fitting catheter combinations [2], the withdrawal of microcatheters through rotating hemostatic valves [7,13], and friction in coaxial catheter systems due to tortuous vascular anatomy [12] have all been reported as mechanisms potentially causing coating damage. Recently, coil delivery during aneurysm treatment has been shown to damage the inner lining of microcatheters, which may contribute to thrombus formation within the microcatheter [23]. The effect of coating impairment on thrombogenicity and navigation capabilities of microguidewires remains to be determined.

Richter et al. [3] reported an increased incidence of NICE lesions associated with certain families of flow diverters in a retrospective, multicenter study involving 1,201 patients treated by flow diversion. However, insufficient data on catheters and guidewires did not allow for identification of the exact cause of this association. Rather than the flow diverter itself, it could be that a particular guiding catheter, microcatheter, guidewire, or microguidewire (or a specific combination of these devices) was responsible for causing NICE lesions. Our study emphasizes that data on all types of endovascular devices, including guidewires and microguidewires (often multiple per procedure), should be collected in future studies to better determine the relative contribution of each device to the risk of NICE lesions.

Limitations

This preliminary study has several limitations. The primary goal of our report was to determine whether the J-wiring technique causes damage to guidewire coatings. Due to our small sample size, we did not aim to assess the variability in coating damage for each guidewire, which could potentially differ between manufacturing batches, or to compare coating damage among different guidewire types. Although the in vitro experiment was carefully conducted, several limitations should be acknowledged: (1) J-shaping and the J-wiring technique were performed manually, which can cause slight variations with each procedure; (2) only one side of the guidewire was imaged using SEM because of the mounting method; (3) high-resolution SEM imaging of the entire guidewire tip and detailed segmentation of coating damage with specialized software would yield a more accurate assessment of coating damage and facilitate better comparisons between guidewires; and (4) due to costs and availability, this study does not include all currently used microguidewires, e.g., the Asahi Chikai (Asahi Intecc) or Aristotle® (Scientia Vascular), and only includes a single flow diverter, which limits its broader applicability. Nonetheless, based on this work and previous studies, it is reasonable to state that mechanical damage to polymer coatings is a common issue affecting most microguidewires [15,19]. Despite these limitations, our study offers valuable insights and underscores the need for more rigorous testing of the durability of polymer coatings on microguidewires used in neurointerventional procedures.

CONCLUSION

Damage to polymer coatings on guidewire tips can result from J-shaping and the J-wiring technique used to optimize flow diverter wall apposition. In our study, both steps damaged the outer and inner curves of the J-loop, with the most extensive changes seen after the J-wiring technique. Differences in the overall patterns of coating damage among guidewires were primarily determined by the guidewire tip design. These processes may generate polymer micro-fragments, increasing the risk of cerebral polymer emboli and the development of NICE lesions in patients treated for cerebral aneurysms with flow diverters. Damage caused by manual shaping of a guidewire tip might pose a risk of introducing micro-fragments into the cerebral circulation during insertion. Using pre-shaped guidewires or thoroughly flushing the guidewire after manual shaping may help reduce this risk. Future studies should examine a wider variety of guidewires under standardized conditions to compare their durability.

Notes

Acknowledgments

We acknowledge the Core Facility for Integrated Microscopy, Faculty of Health and Medical Sciences, University of Copenhagen, Denmark, for contributing the scanning electron micrographs.

Fund

None.

Ethics Statement

Institutional Review Board approval was waived for this study. No patient information was reported in this study.

Conflicts of Interest

GB received research support from and holds patents with Siemens Healthineers. ET, RWL, and RHD have no conflicts to disclose.

Author Contributions

Concept and design: GB, RHD. Analysis and interpretation: GB, RHD. Data collection: GB, RHD. Writing the article: GB, RHD. Critical revision of the article: All authors. Final approval of the article: GB. Overall responsibility: GB.

References

1. Shotar E, Labeyrie MA, Biondi A, Velasco S, Saliou G, Boulouis G, et al. Non-ischemic cerebral enhancing lesions after intracranial aneurysm endovascular repair: a retrospective French national registry. J Neurointerv Surg 2022;14:925–930.
2. Mehta RI, Mehta RI. Polymer-induced central nervous system complications following vascular procedures: spectrum of iatrogenic injuries and review of outcomes. Hum Pathol 2016;53:178–190.
3. Richter C, Möhlenbruch MA, Vollherbst DF, Taschner CA, Elsheikh S, Klisch J, et al. Non-ischemic cerebral enhancing (NICE) lesions after flow diversion for intracranial aneurysms: a multicenter study. J Neurointerv Surg 2024;16:1174–1180.
4. Bayas A, Christ M, Berlis A, Naumann M, Ertl M, Joachimski F, et al. Incidence, clinical spectrum, and immunotherapy of non-ischemic cerebral enhancing lesions after endovascular therapy. Ther Adv Neurol Disord 2022;15:17562864211072372.
5. Ikemura A, Ishibashi T, Otani K, Yuki I, Kodama T, Kan I, et al. Delayed leukoencephalopathy: a rare complication after coiling of cerebral aneurysms. AJNR Am J Neuroradiol 2020;41:286–292.
6. Nakagawa I, Park HS, Kotsugi M, Morisaki Y, Wada T, Aketa S, et al. Delayed intracranial parenchymal changes after aneurysmal coil embolization procedures for unruptured intracranial aneurysms. Oper Neurosurg 2020;19:76–83.
7. Kan I, Karagiozov K, Ito S, Sato S, Murayama Y. Microcatheter originating debris during neuroendovascular procedures: mechanism of dislodgement and its prevention. AJNR Am J Neuroradiol 2020;41:1879–1881.
8. Stanley JR, Tzafriri AR, Regan K, LaRochelle A, Wong G, Zani BG, et al. Particulates from hydrophilic-coated guiding sheaths embolise to the brain. EuroIntervention 2016;11:1435–1441.
9. Mordhorst A, Yang GK, Reitsma N, Chen JC, Zhang BY, Suri S, et al. Evaluation of hydrophilic polymer embolization from endovascular sheath devices in an in vitro perfusion system. JVS Vasc Sci 2023;4:100127.
10. Hu YC, Deshmukh VR, Albuquerque FC, Fiorella D, Nixon RR, Heck DV, et al. Histopathological assessment of fatal ipsilateral intraparenchymal hemorrhages after the treatment of supraclinoid aneurysms with the Pipeline Embolization Device. J Neurosurg 2014;120:365–374.
11. Barnwell SL, D'Agostino AN, Shapiro SL, Nesbit GM, Kellogg JX. Foreign bodies in small arteries after use of an infusion microcatheter. AJNR Am J Neuroradiol 1997;18:1886–1889.
12. Oh SW, Shin NY, Lee HJ, Kim BM, Kim DJ. Delayed enhancing lesions after coil embolization of aneurysms: clinical experience and benchtop analyses. J Neurointerv Surg 2017;9:1243–1247.
13. Müller SJ, Albiña-Palmarola P, Konieczny S, Manke G, Fischer S, Henkes H. Hydrophilic polymer coating delamination during neurointerventional treatment after microcatheter withdrawal: particulate identification through attenuated total reflection Fourier-transform infrared spectroscopy. Front Neurol 2025;15:1479375.
14. Kühn AL, Rodrigues KM, Wakhloo AK, Puri AS. Endovascular techniques for achievement of better flow diverter wall apposition. Interv Neuroradiol 2019;25:344–347.
15. Dahl RH, Larsen RW, Thormann E, Benndorf G. Fragmentation of hydrophilic guidewire coatings during neuroendovascular therapy. Clin Neuroradiol 2023;33:793–799.
16. Chopra AM, Mehta M, Bismuth J, Shapiro M, Fishbein MC, Bridges AG, et al. Polymer coating embolism from intravascular medical devices - a clinical literature review. Cardiovasc Pathol 2017;30:45–54.
17. van den Bosch M, Schreurs P, Geers M. Identification and characterization of delamination in polymer coated metal sheet. J Mech Phys Solids 2008;56:3259–3276.
18. Mehta RI, Rai AT, Vos JA, Solis OE, Mehta RI. Intrathrombus polymer coating deposition: a pilot study of 91 patients undergoing endovascular therapy for acute large vessel stroke. Part I: histologic frequency. J Neurointerv Surg 2019;11:1191–1196.
19. Dahl RH, Larsen RW, Thormann E, Benndorf G. Polytetrafluoroethylene coating fragments during neuroendovascular therapy: an analysis of two damaged microguidewires. Interv Neuroradiol 2022;28:16–21.
20. Liu J, Qu S, Suo Z, Yang W. Functional hydrogel coatings. Natl Sci Rev 2020;8:nwaa254.
21. Peng S, Zhang L, Xie G, Guo Y, Si L, Luo J. Friction and wear behavior of PTFE coatings modified with poly (methyl methacrylate). Compos B Eng 2019;172:316–322.
22. Roy S, Ghosh SK. Influence of medium polyurethane coating on mechanical, wear, and corrosion performance of a galvanised steel sheet. J Alloys Metall Syst 2025;9:100174.
23. Takahashi S, Hirai S, Fujita K, Bongguk K, Kinoshita Y, Wakabayashi H, et al. Scanning electron microscopy analysis of microcatheter inner lining damage and thrombogenesis in coil embolization. Interv Neuroradiol 2025;:15910199251377132.

Article information Continued

Fig. 1.

Radiograph of the silicone vascular model with a PipelineTM Vantage flow diverter (Medtronic) deployed in a curved segment with a 6-mm aneurysm. The J-wiring technique was performed by shaping the tips of the AvigoTM (Medtronic) (shown), Synchro® (Stryker), and TraxcessTM (Terumo Neuro) guidewires into 180° J-loops and moving the J-loop back and forth 3 times inside the flow diverter. The curved arrow indicates the back-and-forth movement of the guidewire tips during the J-wiring technique. The guidewires were moved between the proximal and distal ends (full length) of the flow diverter. Guidewire tips were cut immediately after withdrawal from the model and subsequently imaged by scanning electron microscopy.

Fig. 2.

Scanning electron micrographs of the AvigoTM (Medtronic) guidewire tip. (A, B) Unshaped control: intact coating with a few small irregularities. (C, D) J-shaping alone: continuous coating damage on both the inner and outer curves of the guidewire (arrows). (E, F) The J-wiring technique: continuous coating damage on both the inner and outer curves of the guidewire (arrows).

Fig. 3.

Scanning electron micrographs of the Synchro® (Stryker) guidewire tip (polymer-coated hypotube). (A, B) Unshaped control: the coating is intact. The guidewire was also intact after J-shaping alone (not shown). (C–F) The J-wiring technique: slight coating damage were found on both the outer (arrows in C and D) and inner (not shown) curves of the guidewire on multiple consecutive levels of the hypotube, exposing an irregular surface. A partially attached micro-fragment probably coming off the deeper irregular surface (arrowheads in D). Other types of coating damage that spanned multiple gaps in the hypotube include long, linear scratches (arrows in E) and focal areas of coating being dragged along the hypotube (arrows in F).

Fig. 4.

Scanning electron micrographs of the TraxcessTM (Terumo Neuro) guidewire tip (polymer-coated coiled wire). (A, B) Unshaped control: the guidewire is intact apart from a single focal area of coating damage (arrows). A few irregularities in the coating in the gaps between the coiled wire are observed. (C, D) J-shaping alone: coating damage on both the inner and outer curves of the guidewire (arrows) on multiple consecutive levels of the coiled wire. (E, F) The J-wiring technique: coating damage on both the inner and outer curves of the guidewire (arrows) on multiple consecutive levels of the coiled wire, most extensive on the outer curve. A close-up of the coating damage in (F) shows multiple longitudinal grooves in the coiled wire. The coating damage mainly involves the central part of the coiled wire, while the coating in the gaps between the coiled wire remains intact.