| Neurointervention > Volume 21(3); 2026 > Article |
|
Ethics Statement
This single-center retrospective study was reviewed and approved by the Mayo Clinic Institutional Review Board (IRB no. 26-000165) and waived informed consent because of the retrospective design and use of deidentified data. We anonymized all patient information that could potentially identify an individual.
Conflicts of Interest
HC serves as the editor of the Neurointervention. WB reports consulting fees from Medtronic, Stryker, Imperative Care, MicroVention, MIVI Neurovascular, Cerenovus, Asahi, and Balt; royalties from Medtronic and Balloon Guide Catheter Technology; equity in Nested Knowledge, Superior Medical Editors, Piraeus Medical, Sonorous Medical, and MIVI Neurovascular; and leadership or fiduciary roles with MIVI Neurovascular, Piraeus Medical, and WFITN. GL reports equity in Piraeus Medical. All other authors report no relevant conflicts of interest. All study devices were obtained through routine institutional purchasing; Piraeus Medical did not supply devices, provide study funding, participate in study design, data collection, data analysis or interpretation, manuscript preparation, or the decision to submit or publish the study.
Author Contributions
Concept and design: GEH, DD, WB, GL, HC, and LR. Analysis and interpretation: GEH, DD, JT, CG, WB, GL, and HC. Data collection: GEH, GL, HC, and LR. Writing the article: GEH, DD, JT, CG, WB, GL, and HC. Critical revision of the article: GEH, DD, WB, GL, HC, and LR. Final approval of the article: GEH, DD, JT, CG, WB, GL, HC, and LR. Statistical analysis: GEH, DD, JT, and GL. Obtained funding: none. Overall responsibility: GEH, DD, and WB.
Values are presented as number/total number (%) unless otherwise indicated. Seven patients accounted for 9 access-site component events because 2 patients met more than 1 category. The access-site composite included hematoma, minor bleeding, and pseudoaneurysm; radial artery spasm and radial artery occlusion were analyzed separately. All 6 hematomas were low-grade (EASY I–II), and all access-site events were managed conservatively. Radial artery patency was formally assessed in 15 patients during a period of routine Doppler surveillance, and postprocedural computed tomography was obtained in 51. Transient and persistent deficits are reported among the 4 patients with a new neurologic deficit.
EASY, Early Discharge After Transradial Stenting of Coronary Arteries; CT, computed tomography; mRS, modified Rankin Scale; IQR, interquartile range.
| Device (manufacturer) | Outer diameter | Inner diameter | ID/OD ratio (%)* | Available / reported length (s, cm) | Published success rate (%)‡ |
|---|---|---|---|---|---|
| 87 NeuGlide / NeuGlide+(Piraeus Medical) | 0.100˝ max (≈7.6F) | 0.087˝ | 87.0 | 100–110† | 97.5 [10] |
| RIST 079 (Medtronic) | 0.093˝ (7F) | 0.079˝ | 84.9 | 95, 100, 105 | 96.1 [21] |
| BENCHMARK BMX81 (Penumbra) | 0.097˝ (≈7.4F) | 0.081˝ | 83.5 | 95, 105, 115 | 96.3 [23] |
| Armadillo (Q’Apel Medical) | 0.095˝ (≈7.2F) | 0.072˝ | 75.8 | 95, 105 | 98.0 [20] |
| Zoom RDL (Imperative Care) | 0.110˝ (≈8.4F) | 0.088˝ | 80.0 | 95, 103, 105, 110 | 89.7 [22] |
Published rates are descriptive and not directly comparable because endpoint definitions differ.
ID/OD ratio, inner diameter divided by outer diameter; FDA, US Food and Drug Administration.
* ID/OD ratio calculated as inner diameter divided by outer diameter and is provided as a dimensional comparison rather than a manufacturer-reported performance metric.
‡ Published success rates are taken from the cited source reports. [10,20-23] Patient populations, procedures and definitions of technical success differ across studies; percentages are descriptive and not directly comparable. The 97.5% NeuGlide value is from a prior 40-patient mixed-neuroendovascular series, not from the present 74-patient aneurysm cohort.

![]() |
![]() |