Transradial Intracranial Aneurysm Embolization Using a 7F Sheath-Compatible 0.087-Inch Guide Catheter: A Single-Center Retrospective Study
Article information
Abstract
Purpose
Transradial intracranial aneurysm embolization is well established, but experience with large-bore 7F sheath-based guide platforms remains comparatively limited. We evaluated the technical feasibility and periprocedural outcomes of a 7F sheath-compatible 0.087-inch guide catheter for transradial intracranial aneurysm embolization.
Materials and Methods
In this single-arm, noncomparative retrospective study, we reviewed consecutive patients who underwent attempted transradial intracranial aneurysm embolization using the NeuGlide as the primary guide catheter at a single center between January 2025 and May 2026. Preaccess radial ultrasonography was performed in all patients, and radial diameter ≥1.7 mm was required for 7F access. The primary outcome was completion of the intended treatment through radial access without femoral conversion.
Results
Target-vessel catheterization was achieved in 71/74 procedures (95.9%), and radial procedural completion in 69/74 (93.2%; 95% confidence interval, 85.1–97.1%). Three procedures required femoral conversion, and 2 were terminated without aneurysm treatment; overall treatment completion was 72/74 (97.3%). Guide instability occurred in 2 procedures: 1 repositioning-only event and 1 event resulting in femoral conversion; no guide exchanges occurred. Seven patients had 9 access-site component events, all managed conservatively; all 6 hematomas were low-grade (EASY I–II). Radial artery occlusion was identified in 1/15 patients formally assessed.
Conclusion
In this radial-first cohort, the 7F sheath-compatible 0.087-inch NeuGlide guide catheter achieved a 93.2% radial procedural completion rate across multiple intracranial aneurysm treatment strategies. These findings support the NeuGlide as a technically feasible and versatile large-bore transradial platform in appropriately selected patients.
INTRODUCTION
Intracranial aneurysms affect approximately 3% of adults, and rupture carries substantial morbidity and mortality [1,2]. Endovascular treatment includes coiling, stent-assisted coiling, balloon remodeling, flow diversion, and intrasaccular devices [3]. These procedures require stable guide support, distal access, controlled device delivery, and adequate angiographic assessment.
Transfemoral access remains common because it accomneurointervention modates large guide systems, but groin hematoma, pseudoaneurysm, retroperitoneal hemorrhage, and unfavorable iliofemoral or arch anatomy can complicate its use [4-6]. Transradial access can reduce access-site morbidity, allow earlier mobilization, and provide an alternative in difficult femoral or arch anatomy [4,7]. Its smaller caliber may constrain guide size, support, and adjunctive-device compatibility [8].
Transradial aneurysm embolization has shown high procedural completion rates [7]. Most published experience, however, uses smaller 6F guide systems, and data on 7F sheathbased platforms with larger working lumens remain limited. The Piraeus 7F 87 NeuGlide (Piraeus Medical) is compatible with a 7F radial sheath and has a 0.087-inch inner diameter and 0.100-inch maximum outer diameter [9]. A prior 40-patient series described its use across several neuroendovascular procedures [10]. We evaluated its technical feasibility, device compatibility, and periprocedural outcomes in transradial intracranial aneurysm embolization.
MATERIALS AND METHODS
Study Design and Patient Population
We conducted a single-arm, noncomparative, single-center retrospective study of consecutive patients in whom intracranial aneurysm embolization was initially attempted through transradial access at a tertiary academic center from January 2025 through May 2026. Cases were identified from the institutional neurointerventional procedural log/database. During this period, 111 intracranial aneurysm embolization procedures were performed through all access routes; 74 (66.7%) were initially attempted through transradial access using the NeuGlide and comprised the present cohort. The institution followed a radial-first neuroendovascular practice.
The cohort included ruptured and unruptured, previously treated, saccular, fusiform, blister, and bifurcation aneurysms undergoing attempted primary coiling, stent-assisted coiling, flow diversion, or Woven EndoBridge (WEB) placement. No prespecified aneurysm size, location, morphology, rupture-status, or treatment-technique criteria determined access selection. Each patient contributed 1 procedure and 1 target aneurysm. Inclusion was based on the initial transradial treatment attempt; procedures requiring femoral conversion or termination without aneurysm treatment remained in the primary-outcome denominator.
Seventeen of 74 procedures (23.0%) were included in a previously published 40-patient series evaluating the catheter across multiple neuroendovascular procedure types; all 17 overlapping cases were aneurysm embolizations [10]. The present study includes 57 additional patients. The study was conducted in accordance with institutional requirements and the Declaration of Helsinki [11].
Device and Procedural Technique
The 87 NeuGlide guide catheter is a single-lumen catheter constructed with metal coil/braid reinforcement and medical-grade polymers, with a hydrophilic coating, soft distal tip, and radiopaque distal marker. It has a 0.087-inch inner diameter, maximum outer diameter of 0.100 inches, and is available in 100- and 110-cm working lengths [9]. The 87 NeuGlide+ is the later-generation model, retaining a 0.087-inch inner diameter and 0.100-inch maximum outer diameter while incorporating a linerless construction. NeuGlide+ became available later in the study period and was used in 2 procedures because it was the available model at that time, rather than because of prespecified patient, anatomic, or procedural criteria. Given the identical relevant lumen and outer-profile dimensions and the small number of NeuGlide+ cases, outcomes were pooled. Four neurointerventionists performed the procedures.
Radial artery patency and caliber were assessed by ultrasonography before access in all 74 patients. Transradial access with a 7F sheath was not pursued when radial artery diameter was <1.7 mm; mean radial artery diameter in the included cohort was 2.1 mm. All procedures used conventional proximal radial access. A 7F radial-specific short sheath was used in every procedure; the institutional technique for this platform used a 16-cm 7Fr Prelude IDeal sheath (Merit Medical) or a 7Fr GlideSheath Slender (Terumo). Preprocedural computed tomography angiography (CTA) was not routinely obtained or reviewed solely to determine radial versus femoral access, and no prespecified measurements of aortic arch, great-vessel takeoff, or cervical vascular angulation were applied. A known aberrant right subclavian artery was considered a contraindication to right transradial access. When available from prior or procedural imaging, marked subclavian, brachiocephalic, or cervical vascular tortuosity and anticipated difficulty achieving stable guide positioning were considered in access planning.
The procedural radial-access protocol was consistent with the institutional technique previously reported for this platform [10]. For procedures requiring adjunctive stents or flow diversion, patients received aspirin 325 mg/day plus clopidogrel 75 mg/day before treatment. VerifyNow P2Y12 testing was used with a target P2Y12 reaction unit (PRU) range of 60–240, with <240 PRU serving as the upper treatment threshold. Dual antiplatelet therapy was continued for 3 months, after which clopidogrel was discontinued and aspirin was continued indefinitely. After sheath placement, 5 mg of intra-arterial verapamil was administered through the sheath and flushed with 10 mL of saline, followed by 6,000 U of intravenous unfractionated heparin; the activated clotting time target was 250–300 seconds. The NeuGlide was advanced over a 5F selective catheter and stiff angled guidewire into the intended cervical target vessel; distal access catheters and treatment devices were selected according to anatomy and the planned treatment. A TR Band (Terumo) compression device was used for radial hemostasis, with patent hemostasis routinely attempted. Air removal began 2 hours after sheath removal and TR Band application, followed by removal of 2–3 mL of air every 10–15 minutes until the device was removed. Subsequent clinical documentation was reviewed for delayed access-site and neurologic events.
Control angiography was performed through the NeuGlide, with or without a distal access catheter. Angiography was considered diagnostic quality when it demonstrated the parent artery, aneurysm, treatment construct, relevant branch vessels, distal perfusion, and potential complications. This outcome was evaluated only when the NeuGlide catheterized the target vessel transradially. Fig. 1 illustrates diagnostic angiography and flow-diverter deployment performed through the NeuGlide.
Transradial flow diversion of an unruptured right paraclinoid internal carotid artery (ICA) aneurysm performed through the 7F 0.087-inch NeuGlide guide catheter (Piraeus Medical). (A) Fluoroscopic image showing the NeuGlide navigated from right radial access into the cervical right ICA over a 5F Simmons 2 catheter and stiff Glidewire. (B, C) Frontal and lateral diagnostic right ICA angiograms obtained through the NeuGlide show an 8.1×6.1 mm unruptured ventral paraclinoid aneurysm projecting posteriorly and slightly inferiorly. (D, E) Frontal and lateral unsubtracted images show the deployed 4×16 mm Pipeline embolization device spanning the aneurysm neck, with the distal end at the ICA bifurcation and proximal end in the horizontal segment. (F) Lateral control right ICA angiogram shows marked intra-aneurysmal contrast stasis after device deployment, preserved filling of the ophthalmic, posterior communicating, and anterior choroidal arteries, and no device narrowing.
Data Collection and Outcome Definitions
Demographic, clinical, anatomic, procedural, angiographic, and safety data were abstracted from the electronic medical record, procedure reports, procedural imaging, and institutional databases. Available imaging and procedure reports were reviewed for aortic arch configuration, great-vessel variants, and cervical vascular tortuosity when relevant to catheterization failure, guide instability, femoral conversion, or procedural termination.
The primary feasibility outcome was radial procedural completion, defined as deployment of the intended treatment construct through radial access without femoral conversion. Target-vessel catheterization was defined as NeuGlide placement in the intended cervical vessel at a position sufficient to proceed with treatment. Overall treatment completion included intended treatment through either radial or femoral access. Guide instability was defined as loss of catheter position or support requiring repositioning, exchange, or access conversion. Device compatibility was defined as use of the intended distal access catheter or treatment device without abandonment of the guide platform because of dimensional incompatibility. Femoral conversion was defined as abandonment of the radial treatment route followed by continuation through femoral access.
Two investigators independently reviewed procedure reports and available angiographic images for all procedures and adjudicated procedural and angiographic outcomes. Neither investigator was the treating operator for the cases adjudicated. Diagnostic-quality angiography was assessed independently using the prespecified imaging criteria above; disagreements were resolved by consensus. Procedures requiring femoral conversion or termination without treatment underwent additional review to classify the failure mode. Failure modes were categorized as failure of target-vessel catheterization, loss of guide position, insufficient support, inability to advance or deploy the intended device, intraprocedural complication, or change in treatment strategy.
Periprocedural outcomes included access-site bleeding or hematoma, pseudoaneurysm, radial artery spasm or occlusion, difficult catheter or sheath withdrawal, dissection or perforation, aneurysm rupture, thromboembolic or hemorrhagic complications, vessel or branch occlusion, new neurologic deficits, and adverse events requiring treatment. The access-site complication composite included hematoma, minor bleeding, and pseudoaneurysm; radial artery spasm and radial artery occlusion were analyzed separately. Hematomas were retrospectively graded using the EASY (Early Discharge After Transradial Stenting of Coronary Arteries) scale, with grades I–II considered low-grade/minor. Events occurring during the procedure or before discharge were classified as periprocedural. Formal radial patency assessment was performed in 15 patients during a period in which postprocedural Doppler assessment was incorporated into routine workflow rather than in response to suspected complications. Available post-discharge clinical documentation was also reviewed for delayed access-site and neurologic events.
Angiographic results were classified as complete occlusion, near-complete occlusion or residual neck, or residual filling. Raymond–Roy and O’Kelly–Marotta grades were used when applicable.12,13 Adequate occlusion was defined as complete occlusion, near-complete occlusion, or residual neck. Follow-up imaging was defined as imaging obtained at least 30 days after treatment. Clinical outcomes included neurologic status, modified Rankin Scale (mRS) score, length of stay, and available follow-up.
Statistical Analysis
Analyses were performed with R version 4.5.2 (R Foundation for Statistical Computing). Continuous variables were summarized as mean±standard deviation or median [interquartile range, IQR], as appropriate, and categorical variables as number and percentage. Wilson 95% confidence intervals (CIs) were calculated for the primary outcome and selected proportions. Missing data were analyzed using available-case denominators without imputation. No formal comparative hypothesis testing was performed for the primary technical outcomes. Exploratory comparisons by rupture status used 2-sided Fisher exact tests for binary categorical variables and Fisher–Freeman–Halton exact tests for multilevel categorical distributions. Because only 7 aneurysms were ruptured and no correction for multiple comparisons was applied, these subgroup P-values were considered hypothesis-generating. Reporting followed the STROBE recommendations. As a sensitivity analysis, transradial target-vessel catheterization and radial procedural completion were recalculated, with Wilson 95% CIs, after excluding the 17 patients included in the previously published series.
RESULTS
Patient, Aneurysm, and Procedural Characteristics
During the study period, 111 intracranial aneurysm embolization procedures were performed at the institution, of which 74 (66.7%) were initially attempted through transradial access and comprised the study cohort. Mean age was 59.6±11.1 years, and 52 patients (70.3%) were women. Sixty-seven aneurysms were unruptured, and 7 were ruptured. Mean maximum aneurysm diameter was 6.2±2.6 mm; 70 aneurysms were saccular, 2 were fusiform, and 2 were blister aneurysms. Ten aneurysms had been treated previously (Table 1).
The NeuGlide was the primary guide catheter in all procedures, including the NeuGlide+ configuration in 2 (2.7%), and a 7F radial sheath was used in every case. Final procedural access was right radial in 70 procedures, left radial in 1, and femoral after conversion in 3. Target vessels were the right carotid artery in 31 procedures, left carotid artery in 31, and vertebral artery in 12. The intended treatment strategy was WEB placement in 27 procedures, flow diversion in 25, primary coiling in 20, and stent-assisted coiling in 2. A 5F Sofia distal access catheter was used in 46 procedures (Table 2).
Technical and Access Outcomes
Transradial target-vessel catheterization was achieved in 71/74 procedures (95.9%; 95% CI, 88.7–98.6%). Catheterization failed because an aberrant right subclavian artery prevented guide advancement into the target internal carotid artery, the catheter system herniated into the aorta during internal carotid artery catheterization, or radial artery tortuosity prevented guide advancement. All 3 procedures were subsequently completed after femoral conversion.
Radial procedural completion was achieved in 69/74 procedures (93.2%; 95% CI, 85.1–97.1%). Three procedures required femoral conversion, and 2 were terminated without aneurysm treatment. All 3 converted procedures were completed, yielding overall treatment completion in 72/74 procedures (97.3%; 95% CI, 90.7–99.3%). All 5 procedures not completed through radial access occurred in patients with unruptured aneurysms. By target vessel, transradial target-vessel catheterization was achieved in 29/31 right carotid procedures (93.5%), 30/31 left carotid procedures (96.8%), and 12/12 vertebral procedures (100%); corresponding radial procedural completion rates were 28/31 (90.3%), 29/31 (93.5%), and 12/12 (100%), respectively. Of the 3 femoral conversions, 2 involved right carotid targets because of an aberrant right subclavian artery or radial artery tortuosity, and 1 involved a left carotid target after catheter-system herniation into the aorta. The 2 procedures terminated without aneurysm treatment occurred after successful target-vessel catheterization: 1 right carotid procedure in which the WEB device could not be conformed to a previously clipped aneurysm and 1 left carotid procedure complicated by coil migration requiring retrieval. No vertebral-target procedure required conversion or termination (Supplementary Table 1). Individual non-completion mechanisms are detailed in Supplementary Table 2. No procedure required abandonment of the guide platform because of dimensional incompatibility. Among the 57 patients not included in the prior series, transradial target-vessel catheterization was achieved in 55/57 (96.5%; 95% CI, 88.1–99.0%) and radial procedural completion in 54/57 (94.7%; 95% CI, 85.6–98.2%), consistent with the full cohort (95.9% and 93.2%, respectively).
Diagnostic-quality angiography was obtained in all 71 procedures with transradial target-vessel catheterization. Final angiography demonstrated preserved parent-vessel and branch patency and distal perfusion in all procedures. Guide instability occurred in 2 procedures (2.7%): 1 repositioning- only event and 1 episode of catheter-system herniation into the aorta that resulted in femoral conversion. No guide exchanges were required.
Periprocedural and Clinical Outcomes
Radial artery spasm occurred in 6 patients, including 2 in whom it complicated catheter or sheath withdrawal; both devices were removed after intra-arterial verapamil administration. Seven patients experienced 9 component access-site events: 6 hematomas, 2 minor bleeding events, and 1 pseudoaneurysm. Two patients met more than 1 component category. All 6 hematomas were low-grade (EASY I–II), and all access-site events were managed conservatively. Radial artery spasm and radial artery occlusion were not included in the access-site composite. Radial artery occlusion was identified in 1/15 patients who underwent formal patency assessment (Table 3).
Device migration occurred in 2 procedures, and rescue maneuvers were required in 4. One patient with a ruptured aneurysm had intraprocedural rerupture. Four patients developed new neurologic deficits, 3 of which were transient. The persistent deficit occurred in a patient with subarachnoid hemorrhage complicated by hydrocephalus. No clinically apparent or angiographically recognized thromboembolic events were documented. Postprocedural computed tomography was obtained in 51 patients and demonstrated hemorrhage only in the patient with rerupture (Table 3).
Immediate angiographic grading was available in 55 patients: 7 had complete occlusion, 11 had near-complete occlusion or a residual neck, and 37 had residual filling. Follow-up imaging was available in 32 patients at a median of 8.5 months; 18 had complete occlusion, and 21 had adequate occlusion. Twenty-five of 32 (78.1%) follow-up studies were noninvasive magnetic resonance angiography or CTA (Supplementary Table 3). Exploratory angiographic and discharge outcomes stratified by rupture status, with unadjusted P-values, are provided in Supplementary Table 4.
At discharge, 73/74 patients had mRS score of 0–1, and median length of stay was 1 day (IQR, 1–1). Clinical follow-up was documented in 46 patients. Follow-up mRS was available in 30 patients and was 0–1 in 28. Review of available post-discharge clinical documentation identified no delayed neurologic events among the 46 patients with follow-up; post-discharge vascular assessment was not protocolized (Table 3).
DISCUSSION
This radial-first series achieved 95.9% target-vessel catheterization and 93.2% radial procedural completion, with 97.3% overall treatment completion. Diagnostic-quality angiography was obtained whenever transradial target-vessel catheterization succeeded, and no procedure required guide-platform abandonment for dimensional incompatibility. Because prior studies already establish transradial aneurysm embolization feasibility [7,14,15], these results are confirmatory rather than evidence of superiority. The contribution of this series is characterization of a 7F sheath-based 0.087-inch platform across contemporary aneurysm treatments; the study does not establish that its larger lumen improves efficiency, safety, or clinical outcomes relative to smaller transradial systems.
Aneurysm embolization may require distal support, multiple microcatheters, or adjunctive devices [16-19]. In this series, diagnostic-quality angiography was maintained through the 0.087-inch platform, including with a 5F Sofia distal access catheter, and no case required guide-platform abandonment for dimensional incompatibility. These findings document compatibility across treatment constructs but do not demonstrate a clinical advantage of the larger lumen.
Prior studies report feasibility of transradial treatment for ruptured and unruptured aneurysms [7,14,15]. Radial-specific guide systems have also been summarized in a recent systematic review [20-23]. Table 4 summarizes selected device dimensions and published success rates. Differences in populations, anatomy, access technique, devices, outcome definitions, and success definitions preclude direct comparisons, and the absence of a contemporaneous transfemoral comparator prevents conclusions regarding comparative efficiency, morbidity, length of stay, or clinical outcomes.
Alternative transradial architectures can provide similar working lumens. The 6F Axcelguide Stiff-J has a 0.088-inch inner diameter and 2.70-mm outer diameter and has been reported for anterior-circulation aneurysm treatment [24]. A 100-procedure sheathless 8F Optimo series used a 0.087-inch lumen; 45% involved aneurysm coiling or flow diversion and technical success was 95% [25]. In 211 unruptured aneurysms, Fuga et al. [26] found lower radial artery occlusion and spasm after propensity matching with sheath-based versus sheathless 8F navigation, without significant differences in procedural success or access-site complications. These data show that lumen size, access profile, and radial vascular effects vary by architecture. The present study did not compare sheathbased, integrated-sheath, or sheathless strategies.
Radial completion was 90.3% for right carotid, 93.5% for left carotid, and 100% for vertebral targets. Guide instability comprised 1 repositioning-only event and 1 femoral conversion after catheter-system herniation, with no guide exchanges; 2 other conversions reflected access or great-vessel anatomy. Access planning should account for radial caliber, great-vessel anatomy, guide position, treatment construct, and device-delivery requirements, with femoral conversion available when radial support is inadequate.
Radial spasm occurred in 6 patients, and 7 patients had 9 access-site component events. Although the 9.5% patient-level composite exceeds the pooled rate in prior transradial aneurysm literature [7], definitions and ascertainment differ: our composite actively captured minor hematoma, bleeding, and pseudoaneurysm, while spasm and radial occlusion were separate. All hematomas were EASY I–II and all access-site events were managed conservatively. Radial occlusion occurred in 1/15 patients assessed during routine Doppler surveillance, precluding reliable cohort-wide estimation. Radial caliber, sheath-to-artery ratio, vasodilators, anticoagulation, compression, and catheter-vessel interaction may influence radial outcomes [8,27]. One ruptured aneurysm reruptured intraprocedurally, and 4 patients developed new neurologic deficits, 3 transient. Because postprocedural imaging was not systematic, silent events may have been missed.
Limitations
The retrospective, single-center design, operator-directed access selection, and absence of transfemoral or alternative transradial comparator groups introduce selection bias and limit generalizability. Although the institution follows a radial-first strategy, patients with radial artery diameter <1.7 mm were not selected for 7F transradial access, and a known aberrant right subclavian artery was considered a contraindication to right transradial access. No prespecified aneurysm- based selection criteria or quantitative thresholds for aortic arch, great-vessel, or cervical vascular angulation were used. In current practice, radial access remains largely a matter of operator judgment, and prospective characterization of the anatomic factors governing suitability would make future work more meaningful. Accordingly, the observed completion rate reflects patients screened and selected for radial access and should not be interpreted as the expected success rate among all patients undergoing intracranial aneurysm embolization. The cohort included heterogeneous aneurysm morphologies and treatment techniques. Exploratory comparisons by rupture status are limited by the small ruptured-aneurysm subgroup (n=7), sparse cell counts, and lack of multiplicity correction and should therefore be interpreted as hypothesis-generating.
Seventeen procedures overlapped with a previously published mixed neuroendovascular series [10]. Sensitivity analysis restricted to the 57 non-overlapping procedures yielded target-vessel catheterization and radial completion rates similar to the full cohort. Follow-up imaging was available in fewer than half of patients and at variable intervals; clinical follow-up and post-discharge vascular assessment were not protocol-driven. Formal radial patency assessment was limited to 15 patients despite being performed during a period of routine Doppler surveillance. Patient-level sheath model was not retained as a structured variable across the full cohort, precluding reliable calculation of individual sheath-to-artery ratios. Postprocedural imaging was not systematic, so delayed radial occlusion, minor access-site events, and asymptomatic neurologic complications may have been missed. During the study period, 74 of 111 institutional aneurysm embolization procedures (66.7%) were initially attempted through transradial access, providing context for the selection pathway but not eliminating selection bias. Prospective multicenter studies with prespecified selection criteria, comparator groups, and systematic vascular follow-up are needed.
CONCLUSION
In this radial-first cohort, 93.2% of attempted intracranial aneurysm embolizations were completed through radial access using the 7F sheath-compatible 0.087-inch NeuGlide guide catheter. The NeuGlide demonstrated high procedural completion and compatibility across multiple aneurysm-treatment strategies, supporting its use as a large-bore transradial guide platform in appropriately selected patients. Comparative studies are warranted to further define its performance relative to transfemoral access and other transradial systems.
SUPPLEMENTARY MATERIALS
Supplementary materials related to this article can be found online at https://doi.org/10.5469/neuroint.2026.01032.
Technical outcomes stratified by target vessel
Procedures not completed through transradial access
Angiographic outcomes overall and by treatment technique
Exploratory angiographic and discharge outcomes stratified by rupture status
Notes
Fund
None.
Ethics Statement
This single-center retrospective study was reviewed and approved by the Mayo Clinic Institutional Review Board (IRB no. 26-000165) and waived informed consent because of the retrospective design and use of deidentified data. We anonymized all patient information that could potentially identify an individual.
Conflicts of Interest
HC serves as the editor of the Neurointervention. WB reports consulting fees from Medtronic, Stryker, Imperative Care, MicroVention, MIVI Neurovascular, Cerenovus, Asahi, and Balt; royalties from Medtronic and Balloon Guide Catheter Technology; equity in Nested Knowledge, Superior Medical Editors, Piraeus Medical, Sonorous Medical, and MIVI Neurovascular; and leadership or fiduciary roles with MIVI Neurovascular, Piraeus Medical, and WFITN. GL reports equity in Piraeus Medical. All other authors report no relevant conflicts of interest. All study devices were obtained through routine institutional purchasing; Piraeus Medical did not supply devices, provide study funding, participate in study design, data collection, data analysis or interpretation, manuscript preparation, or the decision to submit or publish the study.
Author Contributions
Concept and design: GEH, DD, WB, GL, HC, and LR. Analysis and interpretation: GEH, DD, JT, CG, WB, GL, and HC. Data collection: GEH, GL, HC, and LR. Writing the article: GEH, DD, JT, CG, WB, GL, and HC. Critical revision of the article: GEH, DD, WB, GL, HC, and LR. Final approval of the article: GEH, DD, JT, CG, WB, GL, HC, and LR. Statistical analysis: GEH, DD, JT, and GL. Obtained funding: none. Overall responsibility: GEH, DD, and WB.