INTRODUCTION
Since the publication of large trials demonstrating the safety and efficacy of endovascular coiling embolization in ruptured intracranial aneurysms, it has become standard care at most centers as the initial treatment strategy [
1].
However, with increased use of coils in progressively complex ruptured aneurysm cases, due to the development of balloon-assisted and stent-assisted techniques, an increase in coiling-related complications has also been observed [
2].
Among these, coil prolapse and coil migration constitute a rare but potentially serious complication, since if not correctly managed it can lead to thrombus formation and vessel occlusion resulting in cerebral infarction, even in a delayed fashion [
3]. Its reported incidence rate ranges from 2% to 6%, which has not significantly changed over time, and consequences have ranged from asymptomatic to occlusion of major arteries, potentially resulting in large territory infarcts [
4].
Different endovascular techniques have been described for managing an intraprocedural coil prolapse or migration: stent placement, microwires, the Alligator Retrieval device, Merci device, snare device, stent retrievers (SR), and even microsurgical extraction, with no standard or truly effective strategy described to our knowledge [
5-
10].
The Embolus Retriever with Interlinked Cages (ERIC) device
® (Microvention, Aliso Viejo, CA, USA) (
Fig. 1A,
B) is a new generation stentriever constructed with an interlinked cage technology, designed to have a high intraluminal metal coverage and less vessel wall contact, theoretically allowing for an atraumatic thrombectomy and better thrombus entrapment. Here we describe our experience using the ERIC device (ED) to recapture a migrated coil, after failed attempts with other stentrievers, as well as an
in vitro demonstration of the efficacy of the device and the reasoning for it to be used as a first approach in cases of displaced or migrated coils.
MATERIALS AND METHODS
Our institutional review board approved this study, and the informed consent was waived.
Clinical and radiological data were retrospectively reviewed from our institutional database for patients who underwent an aneurysm embolization procedure in our institution between January 2017 and December 2019. From this cohort, patients who presented coil prolapse or migration during or after endovascular embolization were identified, in whom at least 1 retrieval attempt was performed. Coil prolapse or migration was defined as a coil coming out of the aneurysmal sac after having been released, either partially (prolapse) or completely (migration). Procedural data was collected, which included coil size, devices employed, and number of retrieval attempts.
Additionally, an
in vitro analysis was performed in a tubular, multicurved segment, silicone flow model of the carotid artery, comparing the ED with other well-known stent retrievers on the effectiveness of coil retrieval (
Table 1). For this, up to 2 attempts were made to retrieve a detached coil (Target
® 7 mm×15 cm, 360 Standard; Stryker, Kalamazoo, MI, USA), employing 1 of 3 stent retrievers (
Table 1), either a closed cell SR (Trevo XP ProVue
®; Stryker), a different articulated coil (Embotrap 2
®; Johnson & Johnson, New Brunswick, NJ, USA), or the ERIC device
®. The technique consisted of detaching a different coil inside the flow model for each device, and then trying to retrieve it with the stent retriever, without using an intermediate catheter or pinning the coil.
DISCUSSION
Thromboembolic events represent the most feared complications associated with endovascular treatment of brain aneurysms [
3], and migration of detachable coils during embolization significantly contributes to this risk. In addition, the coil may migrate further into smaller vessels, producing flow limitation, or worse it may stretch and fracture, creating a thrombogenic nidus leading to a parent vessel occlusion or distal shower of microemboli material, which may result in a large territory infarct or even death. Although most case series report a low incidence of 2–6% of coil migration [
4], the potentially devastating consequences associated with this complication require the development of an effective and secure maneuver to retrieve the displaced coil.
Microsnares and the Alligator device
® were specifically developed for the retrieval of foreign bodies from intracranial vessels [
9], but the difficulty and potential complications associated with the use of snares and the production stop of the Alligator have rendered these devices obsolete. On the other hand, use of the Merci retriever [
6] and guidewires [
7] have also been described in small case series, but the unavailability of the first and unpredictability associated with the second are also a limitation for their use.
Most publications regarding the use of stent retrievers are small case series or case reports like ours. Leslie-Mazwi et al. [
5] reported the use of Solitaire and Catch Plus stent retrievers in a series of 14 patients, and although successful in all cases, only in 6 showed the coil retrieved with a single pass, with 1 patient requiring 7 passes. Zhou et al. [
10] presented a case series of 7 patients with dislodged neurovascular devices, 5 of which presented a coil migration that was successfully retrieved with a stentriever, although no mention was made of the device construction, with more emphasis made on the technique of securing the coil between the device and the microcatheter, as a way of ensuring the retrieval.
Kabbani et al. [
11] reports a case of a prematurely detached coil migrated into the left pericallosal artery, in which after previous failed attempts with Microsnares, a Trevo ProVue
® was used to successfully retrieve the coil in 2 passes. Similarly, Nas et al. [
12] presented another case of a partially protruding coil being retrieved with a Solitaire
® device after 3 passes, with no mention of the device construction in its ability to successfully retrieve the coil. Amuluru et al. [
13], on the other hand, discussed the characteristics of stentrievers and the construction and the superiority of these devices over snares or other retrieval technologies, and even proposed that the use of a large Solitaire
® in a small-caliber vessel would theoretically allow multiple struts crossings in a better coil engagement. The authors reflected on the problem of crossing the dislodged coil with a microcatheter and the possibility of displacing it further into smaller vessels, though they did not comment on the possibility of using smaller microcatheters to reduce that problem.
Finally, Nikoubashman et al. [
14] presented a study in an animal model in which the Trevo ProVue stentriever was used to perform 148 extractions, with a success rate of 72%, and they described a retrieval technique in which the coil was trapped within the stent by partially re-sheathing it, thus locking parts of the coil inside, and achieving a 99% success rate with this technique. In our series, we did not try this retrieval technique, but it should perform equally or even better with the ERIC device due to its multicage configuration. In our opinion, in contrast with other tubular design stentrievers, the specific design of the ERIC device could theoretically allow a better coil trapping between any of the cages (
Fig. 4B), increasing the probability of coil retrieval. Also, the 0.017 inch microcatheter profile could minimize the risk of coil distal displacement in comparison with a 0.021 microcatheter during the rescue maneuvers.
On the other hand, Simgen et al. [
15] reports on another animal model in which different stentrievers were used for volume coils retrieval, and the results reported inferiority of ERIC in comparison with the other 3 stent retrievers. Nevertheless, it should be considered that the difference in coil size may play a role in this result, which may not be easily extrapolated to a regular 0.010 coil retrieval.
The ERIC device has been compared with other stentrievers for thrombectomy, showing equivalent angiographic and clinical results to other devices, albeit with a higher rate of complete recanalization (modified thrombolysis in cerebral infarction scale 3), but it has never been described as an alternative for coil retrieval before [
16].
Edwards et al. [
17] reported antiplatelet medication as a useful tool in managing a coil prolapse in some patients as a sufficient way to avoid thromboembolic complications. However, the authors reported specifically on the management of grade I prolapse, not complete coil migration or a high-grade protrusion. On the other hand, the management should not rely solely on the antiplatelet therapy for our cases. Therefore, we used a loading dose of IV aspirin when the complication was observed and then retrieved the coil.
In our series, the use of an ERIC device was associated with a single pass success for recapturing a migrated coil, and its behavior during in vitro testing was superior to other devices, even though the trapping technique was not used, and the stents were retrieved at maximum opening. Another popular segmented device, the EmboTrap, did not provide the same results as the ERIC. It was probably due to a difference in the cage morphology.
This study has inherent limitations. First, it is a small case series with no control population analysis for comparison. Considering the small incidence of these complications, further studies with multicenter data would be needed. Second, the flow-model analysis lacks a full comparison with other stentrievers, specifically the Solitaire stentriever, a popular device even in this setting. However, our goal was to primarily analyze the difference between using a segmented stentriever and a tubular one. For this purpose, we thought that the Trevo stentriever is sufficient to represent the segmented stentriever. Third, the trapping technique, different coils of various sizes, different anatomical sites, and more attempts per device have not been tested, and they will be considered in further research. Finally, despite the promising results of our report and current data, the superiority of ERIC over other stentrievers has not been proven, thus requiring larger case series.
Despite all this, to our knowledge, this is the first time the ERIC device has been described as a useful tool for recapturing a migrated coil, and after an analysis and review of the cases and in vitro models, we consider it to be an good device for coil migration management, since the multicage configuration and low profile (0.017 inch compatible) allows for a more stable coil trapping, even in distal vessels.